# Beta Bionics' iLet recall covers 35,000 pumps on firmware 3.3.11 and 1.6.6; the fix is an update to 3.3.12 or 1.6.7

We asked: Which iLet pumps does the Beta Bionics firmware recall cover, and how do users fix it?

Answered 2026-11-29T02:09:00.000Z.

35,000 pumps in all: 31,303 color iLet Bionic Pancreas pumps on firmware 3.3.11 and 3,697 monochromatic pumps on 1.6.6, distributed nationwide, according to the Food and Drug Administration's recall record Z-0531-2027, posted Nov. 23. Beta Bionics says host software 3.3.12 (color) and 1.6.7 (monochromatic) resolve the problem, and users install it through the iLet phone app.

The record says the pump keeps an internal count of insulin doses that "can reach its maximum value and reset unexpectedly" after about seven to 11 months of use; the pump then stops delivering insulin and, after 30 minutes, shows an "iLet Malfunction (57)" alarm until it is restarted, and the defect "will eventually recur." To check, the record says to open the iLet app's More > More > iLet Bionic Pancreas > About > Host version, or Menu > Settings > About iLet > Host on the pump. The company's customer care line is 1-855-745-3800, option 1. News Without Pictures first reported the recall on Nov. 28.

Sources:
- FDA recall database, Z-0531-2027, iLet Bionic Pancreas (posted Nov. 23, 2026): https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=222744
- News Without Pictures, Nov. 28: Software update can make iLet insulin pumps stop delivering insulin, FDA recall shows: https://newswithoutpictures.com/story/beta-bionics-ilet-firmware?ref=answer

https://newswithoutpictures.com/answer/which-ilet-pumps-does-the-beta-bionics-firmware-recall-cover
Ask your own question: https://newswithoutpictures.com/ask (MCP tool ask_newsroom)

Source: News Without Pictures (newswithoutpictures.com). Free to republish under CC BY-ND 4.0 with credit and a link.
