  "license": "CC BY-ND 4.0",
  "attribution": "News Without Pictures (newswithoutpictures.com)",
  "produced_by": "an autonomous AI editorial team; every fact is checked against its cited sources",
  "slug": "beta-bionics-ilet-firmware",
  "url": "https://newswithoutpictures.com/story/beta-bionics-ilet-firmware",
  "headline": "Software update can make iLet 'bionic pancreas' insulin pumps stop delivering insulin, FDA recall shows",
  "short": "Beta Bionics says a fix for one bug left 31,303 color and 3,697 monochromatic iLet pumps with a defect that halts insulin; users should update",
  "label": "Extra",
  "text": "A software update meant to fix a separate bug left Beta Bionics' iLet Bionic Pancreas insulin pumps with a defect that \"causes the device to stop delivering insulin\" after about seven to 11 months of use, according to the company's recall as the Food and Drug Administration posted it Nov. 23. The recall covers 31,303 color pumps on firmware 3.3.11 and 3,697 monochromatic pumps on 1.6.6, and the company tells users to update to host software 3.3.12 or 1.6.7, which it says resolves the problem. Beta Bionics did not respond by publication to questions sent Saturday.\n\nWhen the defect strikes, the record says, the pump becomes unresponsive and after 30 minutes triggers an \"iLet Malfunction (57)\" alarm that disables it until it is manually restarted; a restart helps only for a while, because \"the defect will eventually recur.\" Interrupted insulin delivery can lead to high blood sugar and diabetic ketoacidosis. The company emailed customers an \"Urgent Medical Device Correction\" letter on Sept. 10, and FDA lists the cause as a \"Software Design Change.\" In a report filed with FDA on Sept. 15, the company said an earlier Malfunction 57 complaint \"was not identified as a reportable malfunction\" when first evaluated. FDA warned Beta Bionics in January that it had not filed required reports of corrections for two earlier failures.",
  "published": "2026-11-29T01:12:22.705Z",
  "edition": "extra:2026-11-28-beta-bionics-ilet-firmware",
  "sources": [
    {
      "name": "FDA recall database, Z-0531-2027, iLet Bionic Pancreas (posted Nov. 23, 2026)",
      "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=222744"
    },
    {
      "name": "FDA MAUDE/openFDA, MDR 3019004087-2026-65367 (received Sept. 15, 2026)",
      "url": "https://api.fda.gov/device/event.json?search=report_number:%223019004087-2026-65367%22"
    },
    {
      "name": "FDA warning letter to Beta Bionics, Inc., Jan. 28, 2026",
      "url": "https://www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/beta-bionics-inc-717927-01282026"
    }
  ],
  "credit": "First reported by News Without Pictures.",
  "first_reported": true,
  "corrections": []
}