  "headline": "Class I recall hits 30,137 NICU and heart kits",
  "short": "Infusion-kit recall",
  "label": null,
  "text": "Cardinal Health is recalling 30,137 Presource kits that contain BD Alaris pump infusion sets, and the Food and Drug Administration has classified the recall Class I. The kits carry labels for NICU admissions, open-heart anesthesia and special procedures, and went to California, Texas and North Carolina.\n\nThe FDA's enforcement report gives the reason in one line: \"Notification of performance disclosures of kit component.\" The full recall record says more. The component is the BD Alaris pump infusion set in each kit, and BD has warned that some of its sets perform worse than it claimed, most often at rates under 1 milliliter an hour. That can mean too much or too little drug, late alarms on a blocked line and inaccurate doses, which BD said \"may not be readily detected by users.\" Newborns are among the patients most at risk. BD's advice is another pump where these sets aren't appropriate, and close monitoring where there is none; it reported no serious injuries or deaths as of Aug. 19. The recall, which Cardinal began voluntarily in September, is still listed as ongoing.",
  "published": "2026-11-28T22:00:00.000Z",
  "edition": "2026-11-28-evening",
  "sources": [
    {
      "name": "FDA medical device recall record Z-0330-2027 (Cardinal Health Presource kits, Class I)",
      "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=223310"
    },
    {
      "name": "Oncology Nursing Society: Cardinal Health updates use instructions for Presource packs containing BD Alaris pump infusion sets",
      "url": "https://www.ons.org/publications-research/voice/news-views/10-2026/cardinal-health-updates-use-instructions-presource-packs-containing-bd-alaris-pump-infusion-sets"
    }
  ],
  "credit": null,
  "first_reported": false,
  "corrections": [
    {
      "date": "Nov. 28, 2026",
      "text": "An earlier version said the FDA's stated reason for the recall ran to eight words. \"Notification of performance disclosures of kit component\" is seven words."
    },
    {
      "date": "Nov. 28, 2026",
      "text": "An earlier version said the FDA's records do not say which kit component is at fault or what it does wrong, and that hospitals got only the one-line reason. The FDA's recall record names the component, the BD Alaris pump infusion sets in the kits, and says Cardinal sent hospitals BD's letter explaining the problem, its clinical impact and what to do. This story now links that record, in place of the enforcement report's search page, and a nursing society's summary of BD's warning.",
      "retired": [
        "same seven words",
        "does not say which component"
      ]
    }
  ]
}