Software update can make iLet ‘bionic pancreas’ insulin pumps stop delivering insulin, FDA recall shows

A software update meant to fix a separate bug left Beta Bionics’ iLet Bionic Pancreas insulin pumps with a defect that “causes the device to stop delivering insulin” after about seven to 11 months of use, according to the company’s recall as the Food and Drug Administration posted it Nov. 23. The recall covers 31,303 color pumps on firmware 3.3.11 and 3,697 monochromatic pumps on 1.6.6, and the company tells users to update to host software 3.3.12 or 1.6.7, which it says resolves the problem. Beta Bionics did not respond by publication to questions sent Saturday.

When the defect strikes, the record says, the pump becomes unresponsive and after 30 minutes triggers an “iLet Malfunction (57)” alarm that disables it until it is manually restarted; a restart helps only for a while, because “the defect will eventually recur.” Interrupted insulin delivery can lead to high blood sugar and diabetic ketoacidosis. The company emailed customers an “Urgent Medical Device Correction” letter on Sept. 10, and FDA lists the cause as a “Software Design Change.” In a report filed with FDA on Sept. 15, the company said an earlier Malfunction 57 complaint “was not identified as a reportable malfunction” when first evaluated. FDA warned Beta Bionics in January that it had not filed required reports of corrections for two earlier failures.

Software update can make iLet 'bionic pancreas' insulin pumps stop delivering insulin, FDA recall shows. First reported by News Without Pictures, November 28, 2026. https://newswithoutpictures.com/story/beta-bionics-ilet-firmware

Free to republish

Know someone who should see this? Email it to them.

Sources

Next · the day’s lead story, 1 of 8Sandia’s Honeywell manager gets $63.55 billion no-bid extension